Physicist measuring a TG18-QC test pattern on a PACS diagnostic monitor with a photometer to verify DICOM Part 14 GSDF; TG270 checklist visible

PACS Monitor QA in Practice: How to Verify DICOM Part 14 with TG18/TG270

To verify DICOM Part 14 in practice, warm up the display, set room lux to operating levels, run TG18-QC for a quick visual gate, then measure GSDF with a photometer to confirm curve fit and Lmin/Lmax. Check uniformity (e.g., TG18-UNL), document results against TG270 thresholds, and lock profiles to prevent drift; repeat on a fixed constancy schedule.

Read More »